More Info... Sensor Simulator

Is there a simulator available for checkout purposes?

Standard System
Sensor Simulator P/N: 186-0105 Instructions For Use:

Description of device

The Sensor Simulator is a service tool that allows for the verification of proper impedance values being detected by the BIS monitors during the Sensor Check. This test is part of the initial test that each monitor performs. The simulator also allows for safety testing of BIS monitors in the field by allowing connection of the test equipment to the monitor via the patient interface cable.

NOTE: The life expectancy of the Sensor Simulator is 25 connect / disconnect cycles.

Schematic of Sensor Simulator circuit

S1 - S4 connect to 4 input signal pins on the patient interface cable. The Inputs (+ and -) are where one connects the test signals to test the BIS Monitor.

Sensor Schematic

Schematic of Sensor Simulator circuit

Test types allowed

Sensor Check:
This checks and verifies that the monitor is reporting the proper impedance that it sees from the Sensor Simulator. This procedure verifies the proper functionality of the BIS monitoring system.
Connect the sensor simulator to the BIS Monitor at the patient interface cable.
The monitor should proceed to recognize that a sensor was connected and report the proper impedance values:

 Electrode # 

 Typical Value in K Ohms 

 Resolution Range in K Ohms 

1

5

4 - 6

2

10

8 - 12

3

2

1 - 3

The monitor then proceeds to display the BIS screen. After a few seconds, BIS = 0 and
SQI = 100.

Safety Testing: Leakage Current.
Leakage Current testing should be performed by a qualified Biomedical Engineering Technician or authorized personnel only.

  • Connect sensor simulator to the patient interface cable of the BIS monitor as if it was a sensor connection.
  • Short the two circular terminals at the end of the simulator using conventional methods such as jumpers or alligator clips. Wire attached with screws will work also.
  • Connect the test signal to the shorted terminals. Make sure that you are not touching the simulator beyond this point.
    Proceed to test instrument for Leakage current as per established facility protocols and procedure for safety testing of medical devices.

Sensor Simulator

Sensor Simulator

Source: A-2000 Service Information Manual P/N 075-0013 Rev.1.01

 

For technical assistance contact:

In the U.S.A.:
Technical Service Department
Aspect Medical Systems, Inc.
Tel: 800-442-2051 (USA)
Tel: 617-559-7000 (USA)
Email: techsupport@aspectms.com

Outside the U.S.A.:
Technical Service
Aspect Medical Systems B.V.
Haagse Schouwweg 8b
NL - 2332 KG Leiden
Tel: +31-715 725 935

<< Back to Frequently Asked Questions | Back to Top

Copyright © 2008. Aspect Medical Systems. All Rights Reserved. Terms of Use | Privacy
Powered by Medigent®

080-0296 1.00